UniVert User Manual
19
Step 7: Execute the Test
Select
Execute
from the File menu or press
on the toolbar.
Step 8: Terminate the Test Prematurely (optional)
You can stop a test at any time by clicking
on the toolbar, or by selecting
Stop
from the File
menu. These are preferred methods of ending a test instead of powering down the equipment or
using the E-Stop, as it allows the software to store the current actuator position.
At the end of the completed test (both normally and prematurely terminated) the actuators will
maintain their current position.
In an emergency, the Emergency Stop (E-Stop) button can be pressed to stop the motor from
moving.
This button will stop power to the UniVert acutator and allow it to be moved by
hand. The displacement position cannot be stored if the E-Stop has been depressed.
Once the UniVert is ready to be used again, twist the E-Stop to return it to its original position. A
new test cannot begin until the E-Stop has returned to its original position. The actuator will need
to be reset to restore the position. To reset the actuator, select
Reset Actuator
from the Tools
menu, or press
on the toolbar. An actuator reset is required before beginning a new test.
Summary of Contents for UniVert
Page 1: ...UniVert Mechanical Test System User Manual version 3 0 ...
Page 36: ...UniVert User Manual 31 ...
Page 78: ...UniVert User Manual 73 ...
Page 79: ...UniVert User Manual 74 ...
Page 80: ...UniVert User Manual 75 ...
Page 81: ...UniVert User Manual 76 ...
Page 82: ...UniVert User Manual 77 ...