C.B.M. S.r.l. a socio unico - Medical Equipment –
Via Castello 10, 26038 Torre dé Picenardi (CR) - Italy
Rev. 06 – 18/02/2019
Manuale SST1700-2200 r06 CE-1282 EN
page 21 of 34
STERILIZING CYCLE
Material Type
Temperature
Time
Drying
Cycle Type S
flash
- Unwrapped solid instruments
and wrapped simple hollow
item
134°C
4 min.
0 / 5 / 10 min.
Cycle Type B
standard
-
Wrapped solid instruments
-
Porous, narrow lumen or
simple hollow item bodies
134°C
4 min.
0 / 15 / 30 min.
Cycle Type B
prion
-
Wrapped solid instruments
-
Porous, narrow lumen or
simple hollow item bodies
134°C
20 min.
0 / 15 / 30 min.
Cycle Type B
delicate
-
Wrapped solid instruments
-
Porous, narrow lumen or
simple hollow item bodies
121°C
18 min.
0 / 15 / 30 min.
LOAD RECOMMENDATIONS
A. Check that the instruments can be steam sterilized.
B. Check that the instruments are
CLEAN
and
DRY.
C. Do not obstruct the tray perforations completely so as to allow the steam to circulate
freely.
D. In case of material packed in paper – plastic bags, put the load in the appliance so that
the paper side is turned downwards;
MAX CHARGE TABLE
SST 1700
SST 2200
Cycle type B
Cycle type S
Cycle type B
Cycle type S
Solid
4,5 Kg
1 Kg. (*)
6 Kg
1,5 Kg (*)
Porous
1,5 Kg.
- - -
2 Kg.
- - -
Remember: the Type S 134°C – 4min. Flash cycle is exclusively dedicated to the
sterilization of wrapped single material or unwrapped or simple hollow item
material.
The cycle
TYPE S
is shorter because the initial vacuum phases are shorter.
Remember that the drying function OFF must not be used with wrapped materials