4. Check that the appliance is positioned slightly inclined towards the back. To this end,
pour a small quantity of water (10cc) in the chamber and ensure it flows towards the
bottom.
Should the inclination need to be changed, regulate the adjustable feet.
WIRING SYSTEM:
Connect the power cable located at the back of the appliance (41),
open the cover below the front display and press the green power switch (13
).
The appliance must be connected to a grounded mains socket in compliance with IEC
Standards, with a mains voltage of 230 Volts and minimum capacity of 16 Amps.
WARNING!!
♦
Do not connect other appliances to the same socket.
♦
Should it be necessary to use an adapter, ensure its capacity is higher than the
maximum sterilizer power ( 2,090 watts).
♦
Failure to observe these precautions may cause malfunctions of the appliance and
hazards for the user.
FIRST USE
1. Provide for a container with at least 3 litres of DEMINERALIZED water, see also the
table at page 24.
WARNING:
the use of ordinary water causes damages and malfunctions of the
appliance in a short time and invalidates its warranty.
2. Switch on the green mains switch (13) located under the cover below the touch-
screen display
3. Fit the appropriate hosepipe (supplied) by pressing the pipe-fitting onto the blue “
Fill
Clean Water Tank”
coupling (12) on the left.
4. Place the other end of the hosepipe in the demineralized water can making sure it
touches the bottom.
5. Press the Right direction key on the display then press the Fill WATER
key.
The tank holds a maximum of 3 litres of water; when the tank is full, FILLING will stop
automatically.
Should water in the container finish before the tank is full, please press the STOP
key on the display.