• Description of the alarms
Alarm code
AC300
DESCRIPTION
SOLUTION
00
0
Microcontroller memory alarm
Contact technician
I14
2
95% amalgam level exceeded
Replace amalgam container as soon as possible
I15
3
100% amalgam level exceeded
Replace amalgam container
I00
32
Microcontroller memory alarm
Contact technician
I01
33
Short-circuit due to one of the two
motors
Check origin of short-circuit and eliminate it
I02
34
Short-circuit before motor control
Contact technician (card probably damaged)
I03
35
Failed capacitor charging
Contact technician (replace card)
I04
36
Temperature threshold exceeded
Ventilate plant room
I05
37
Blower current limit exceeded
Check effi ciency of blower (shutdown or excessive
resistance)
I07
39
Maximum voltage on capacitors
exceeded
Check mains voltage max. 500V
S08
40
Centrifuge short-circuit
Remove the short-circuit
S09
41
Short-circuit of card in centrifuge
outlet
Replace the card
S10
42
Instantaneous current centrifuge
exceeded
Eliminate siphons in pipes or check effi ciency of
centrifuge (shutdown or excessive resistance)
S11
43
Delayed current limit of centrifuge
exceeded
Eliminate siphons in pipes or check effi ciency of
centrifuge (shutdown or excessive resistance)
I16
48
Vacuum sensor pipe disconnected
Connect vacuum pipe to centrifuge
54
Summary of Contents for MAXI SMART
Page 2: ......
Page 30: ...28...
Page 58: ......
Page 86: ...84...
Page 114: ...112...
Page 142: ...140 SCHEMA DI MONTAGGIO MAXI SMART INSTALLATION LAYOUT Fig A Draw A Fig B Draw B...
Page 143: ...141 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 144: ...142 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 145: ...143 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 146: ...144 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 148: ...146 Fig E Draw E DIMENSIONI MAXI SMART MAXI SMART DIMENSIONS...
Page 149: ......
Page 150: ......
Page 151: ......