Special maintenance operations must be carried out by a
trained technician using original spare parts:
• We recommend you pay special attention to all the dan-
ger signs and use protective glasses, gloves and disposa-
ble coverall.
• Check ordinary maintenance effi ciency, ensure that Ma-
gnolia products are used.
• Before intervening on used equipment run a few washing
cycles with Fast & Steril 3
(D)
, wait 15 minutes to allow the
disinfectant to work.
• Check the maximum temperatures recorded and all the
alarm signals, intervene as required.
• Check noise level of suction unit, max 77 dB for units wi-
thout cover and 68 dB with cover, measured in accordance
with regulation 3047 (E).
• Use a jet of dry air, with maximum 2 bar pressure, to clean
any electronic components of the AC300 circuit that need
cleaning.
• Check the plastic pipes, in particular the ones under pres-
sure (at the amalgam separator outlet) which should be re-
placed every 12-18 months;
• Hydrocyclone ISO 60
Check effi ciency with aspirator switched on, suck up a
quantity of Fast & Steril 3 suffi cient to ensure complete di-
sinfection. Remove the amalgam retention vessel and pla-
ce a metal object between the diodes, the aspirator should
read I14 or I15. Finally, check that the holes conveying the
liquid are unobstructed in the Hydrocyclone cone. Before
carrying out any procedure check
the manual for Hydrocyclone ISO 60
• Check effi ciency of the centrifugal separator
(11)
and of
the recirculation valve
(14)
.
.
• Rubber parts: O-rings, bellows, seals should be replaced
every time one of these is removed when a part is disas-
sembled.
• Replace motor bearings.
• If regular maintenance is not performed properly or ina-
dequate products are used, staff should be instructed to
report this to the person in charge in the surgery.
Recommended checks every 12 months
Recommended checks every 18-24 months
Recommended checks every 10.000 months
B
D
A
40
• Extraordinary maintenance
ALLARME:
TEMPERATURA ELEVATA
11
14
Summary of Contents for MAXI SMART
Page 2: ......
Page 30: ...28...
Page 58: ......
Page 86: ...84...
Page 114: ...112...
Page 142: ...140 SCHEMA DI MONTAGGIO MAXI SMART INSTALLATION LAYOUT Fig A Draw A Fig B Draw B...
Page 143: ...141 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 144: ...142 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 145: ...143 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 146: ...144 COLLEGAMENTI CIRCUITO INVERTER AC300 INVERTER AC300 CONNECTIONS Fig C Draw C...
Page 148: ...146 Fig E Draw E DIMENSIONI MAXI SMART MAXI SMART DIMENSIONS...
Page 149: ......
Page 150: ......
Page 151: ......