28
1. General Function
Menu-GEN.01: Set AD Speed
Function
AD Speed Setting
Set Range
(0 ~ 6)
Display Part
Meaning
ADC. 0
Converting speed 5 times per second
ADC. 1
Converting speed 10 times per second
ADC. 2
Converting speed 20 times per second
ADC. 3
Converting speed 40 times per second
ADC. 4
Converting speed 100 times per second
ADC. 5
Converting speed 160 times per second
ADC. 6
Converting speed 320 times per second
Note 1. After changing the AD speed, press ‘
Set' key
to display the weight using the altered value.
(Press ‘
Set'
after checking the changed value to return to the previous mode)
Menu-GEN.02: Set average filter
Function
Average filter Setting
Set Range
(1 ~ 50)
Display Part
Meaning
Fil. XX
Initial Value : 10
Setting the number of buffers in the average
filter
Note 1. Set based on the environment. (The rate of change of weight may slow)
Menu-GEN.03: Set digital filter
Function
Average filter Setting
Set Range
(0 ~ 5)
Display Part
Meaning
IIR. XX
Initial Value : 0
0 : Filter not use
1 ~ 5 : Filter intensity
Note 1. Set based on the environment. (The rate of change of weight may slow)
Menu-GEN.04: Set use key lock
Function
Set use key lock
Set Range
(0 ~ 1)
Display Part
Meaning
LOCK. 0
Unlock front key
LOCK. 1
Lock front key
Menu-GEN.05: Set Key Operating Conditions
Function
Key Operating Conditions(Zero, Tare) Setting
Set Range
(0, 1)
Display Part
Meaning
K-ST. 0
Always in operation
K-ST. 1
Operates only if the weight is stable