26
SET MODE
(1) How to Enter
Turn on the power while pressing the "TARE" key on the front of the indicator.
(2) Available Keys
Used to save inputted value and exit to menu selection.
Used to set the current value to zero.
Used to set the current value ×10.
Used to set the current value +1.
Used to set the current value -1.
(3) Set Value Conversion Menu
Classification
Menu
SubMenu
Initial Value
1. General Setting
(1. GEN)
GEN.01
Set AD speed
1
GEN.02
Set average Filter Buffer
10
GEN.03
Set digital filter buffer
0
GEN.04
Set use key lock
0
GEN.05
Set Key Operating Conditions
1
GEN.06
Set stable range
1
GEN.07
Set Automatic Zero Tracking
Compensation
1
GEN.08
Set Weight Back-up
0
GEN.09
Set Zero range
10
GEN.10
Set Tare range
100
GEN.11
Set Init Zero
10
GEN.12
Set Overload range
9
GEN.13
Set Hold Type
0
GEN.14
Set Hold Canceling Conditions
0
GEN.15
Set average hold time
30
GEN.16
Set Auto Hold Range
10
GEN.17
Set Auto Hold Canceling Range
10
GEN.18
Set the use of Set key
0
GEN.19
Set the use of Print key
6
GEN.20
Set Near Zero(Print, Relay)
0
GEN.21
Initializing of set values
0