CMD
is a two character hexadecimal field:
CMD
Command
Description
05
H
Read Literal
Read register contents in a ‘human readable’ format
11
H
Read Final
Read register contents in a hexadecimal data format
12
H
Write Final
Write the DATA field to the register.
10
H
Execute
Execute function defined by the register using
parameters supplied in the DATA field.
REG
is a four character hexadecimal field:
REG
Register
Description
0005
H
Serial Number
Returns sensor serial number
0020
H
ADC Sample
Number
Read current sample number since last power on. (32
bit)
0021
H
System Status
This register can be read to obtain the status of the
instrument.
32 status bits sent as 8 hex chars, where:
00020000
H
: Overload
00010000
H
: Underload
00008000
H
: Error (see System Error)
00004000
H
: SETUP menus active
00002000
H
: Calibration in progress
00001000
H
: Motion
00000800
H
: Centre of Zero
00000400
H
: Zero
00000200
H
: Net
0022
H
System Error
Diagnostic Errors
0023
H
Absolute mV/V
Absolute mV/V reading where 10000 = 1.0mV/V
0025
H
Displayed Weight Gross or Net weight depending on which is active
0026
H
0027
H
0028
H
Gross,Net,Tare
Weight
Gross Net or Tare weights