37
36
n
Reason
The display weight is larger than the Maximum Capacity you’ve set.
n
Trouble shooting
Don’t load the article whose weight is larger than the Max. Capacity on the platform
scale. This may damage Load Cell.
(2) In CAL Mode
ERR 21
n
Reason
Resolution (Maximum Capacity
÷
Minimum Division) is over the limit (1/10,000).
n
Trouble shooting
Lower the resolution in one of the below ways.
①
Modify Maximum Capacity in CAL 1 of Calibration Menu.
②
Modify Minimum Division in CAL 2 of Calibration Menu.
ERROR MESSAGE & TROUBLE SHOOTING
(1) In Weighing Mode
ERR 02
n
Reason
Load cell connection failure or error in A/D conversion part.
n
Trouble shooting
Check the Load Cell connector to see if the polarity of signal is reversed.
ERR 13
n
Reason
Zero range deviate from the set range.
n
Trouble shooting
Confirm that there is nothing on the weighing platform.
If nothing exist, do Calibration in CAL mode.
ERR 22
n
Reason
Setting weight is set under 10% of the Maximum Capacity.
n
Trouble shooting
Set span setting weight equal to or over 10% of the Maximum Capacity in CAL 3 of
CAL Menu.
OVER
ERR 24
n
Reason
Load Cell output Volltage is too low at span Calibration.
ERR 25
n
Reason
Load Cell output Volltage is too high at span Calibration.