ENGLISH
15
3.5 Input Range Selection
When you have selected the range on the input module (according to
3.4 Jumper Setting on Modules) you have to update or check the pro-
gramming to be sure that the programmed range code corresponds to
the range selected with the jumpers on the module. This is accom-
plished in calibration mode. Other functions in calibration mode are dis-
cussed in chapter 8.1 Calibration Mode. To change or check the pro-
grammed range follow the description below.
1. Press and simultaneously and switch on the unit. The
display shows
. Release the keys. The display shows
for 2 s and then .
2. Press until ‘66’ is displayed. Press and the display now
shows .
3. Press to accept your entry to selection of range code. Now the
display shows the current range code (a number between 0 and 12).
4. Press or to select a range code between 0 and 12
according to the range selected on the module and the list below.
▲
S
▼
S
S
▲
▼
0
All
1
0.2 V
0.2 mA
2
2 V
2 mA
3
20 V
20 mA
4
200 V
0.2 A
5
2 A
6
690 V
5 A
10 ADC
7
0.2 V
0.2 mA
200
Ω
200 Hz
200 rpm/30 ppr
All
8
2 V
2 mA
2 k
Ω
2 kHz
200 rpm/60 ppr
9
20 V
20 mA
20 k
Ω
200 rpm/100 ppr
10
200 V
0.2 A
200 k
Ω
2000 rpm/30 ppr
11
2 A
2000 rpm/60 ppr
12
690 V
5 A
10 AAC
2000 rpm/100 ppr
Range Temp. VDC VAC ADC AAC 10 A Ohm Freq. Tachometer Dup-
code
AC/DC
line
5. When the desired range code is displayed press . The display
shows for 2 seconds and automatically returns to RUN-mode.
The range selection is now completed, and the panel meter is ready to
use. Now you can go ahead with customizing the program.