CS 9300C User Guide (SM750)_Ed 01
6–1
Chapter 6
MAINTENANCE
Perform the following maintenance activities on your CS 9300C unit.
Daily
Cleaning the CS 9300C unit
1. Switch off the unit.
2. Clean all accessible parts of the unit using low-level disinfectants that are U.S.
Environmental Protection Agency (EPA)-registered or low-level disinfectants that are
recognized by your Local Authority. An EPA-registered hospital disinfectant or any
other low-level disinfectant must have clear label claims for intended use. You must
follow the manufacturer instructions for use.
Cleaning and disinfecting the Accessories
Cleaning and disinfecting the accessories that have contact with the
mucous membranes
The accessories are delivered unsterilized. To ensure maximum hygienic safety for the
patient, follow the following instructions carefully to prepare them for use:
The following accessories must be heat-sterilized or disinfected with high-level
disinfectant between each patient use:
TMJx4 nose rest
Panoramic standard bite block
Bite block for edentulous patient
3D bite block
To clean and disinfect the accessories:
1. Remove all visible soil by scrubbing with a surfactant, detergent, and water.
CAUTION
Avoid applying any cleaning liquid to the inside parts of the unit.
CAUTION
We recommend that you cover the accessories with
FDA-cleared protective sheaths that are available from
distributors. After you remove and discard the protective
sheath, each accessory must be cleaned and disinfected
between each patient use.
Summary of Contents for CS 9300C
Page 1: ...CS 9300C User Guide...
Page 6: ...Conventions in this Guide 1 2 About This Guide...
Page 16: ...Positioning Accessories and Replacement Parts 2 10 CS 9300C OVERVIEW...
Page 28: ...Starting the Imaging Software 4 6 GETTING STARTED...
Page 53: ...Acquiring a Submento Vertex Image CS 9300C User Guide SM750 _Ed 01 5 25...
Page 62: ...Annually 6 4 MAINTENANCE...