1000DF00645 Issue 2
10/26
IVAC® 598 Volumetric Pump
Loading an Infusion Set
Loading an Infusion Set
A
Ensure the appropriate infusion set for the fluid/drug to be infused has been selected.
Follow the instructions supplied with the individual infusion set.
Only use IVAC® 598 Volumetric Pump infusion sets, (Refer to 'Infusion Sets' section of the DFU)
Position the fluid container to avoid spillage onto the pump.
Ensure that the tubing is inserted completely into the top set retainer through to the tubing guide avoiding any
slack.
Prime the IV infusion set slowly and close the roller clamp.
1.
Air-In-Line Sensor
Latch
(shown open)
Top set retainer
Primed IV
Infusion Set
Lower Tubing
Segment
Open the door.
2.
Load the primed IV infusion set:
3.
Open the latch
•
Load the upper tubing segment into the top set retainer
•
Insert lower tubing segment into the Air-In-Line Sensor
•
Close the latch, ensuring the set is enclosed
•
Close the door.
4.
Open the roller clamp. Ensure no fluid flows.
5.