1000DF00645 Issue 2
16/26
IVAC® 598 Volumetric Pump
Flow Sensor Operation (Optional)
Flow Sensor Operation (Optional)
A
The flow sensor automatically detects drops falling in the drip chamber. The flow sensor will cause the pump to
alarm if a significant deviation from the infusion rate occurs. The flow sensor will also be able to detect empty
containers. For this reason we recommend use of a flow sensor wherever possible.
IVAC® Flow Sensor Model 192
Handles
Drip chamber
Flanges
Pull back handles
Plug the flow sensor into the flow sensor interface located on the top rear part of the pump.
1.
Attach the IVAC® Flow Sensor Model 192 to the drip chamber of the infusion set, by pulling back the handles. Refer to the illustration
2.
above.
Proceed with load, priming, and set-up instructions as described in section 'Getting Started'.
3.
NOTE:
Ensure drip chamber is half full and upright.
A
Always attach the flow sensor before you start an infusion .
Avoid using the flow sensor in direct sunlight.
Always ensure lens is clean.