1000DF00645 Issue 2
15/26
IVAC® 598 Volumetric Pump
Alarms and Display Messages
Alternating Messages
Display
Cause
Action
bat.
(alternating with rate
display, pump continues
to run)
Battery charge low with
approximately 1 hour of operation
remaining.
Plug the power cord into an AC outlet.
HOLD
(alternating with rate
value)
Pump has been placed on hold. This
is not an alarm condition.
Press
RUN/HOLD
switch once to restart infusion.
Rate
(alternating with
rate value)
Pump is in rate setting mode. This is
not an alarm condition.
Press
RUN/HOLD
or
VOL TO BE INF
switch to advance to VTBI
setting mode.
VtBI
(alternating with
VTBI value)
Pump is in VTBI setting mode. This is not
an alarm condition.
Press
RUN/HOLD
switch once to start infusion.
Prompts
Display
Cause
Action
set vtbi
VTBI value is at zero.
Set the VTBI value to a non-zero value or operate the pump with a
flow sensor and set VTBI to OFF.
Set rate
Rate set to zero.
Enter a non-zero rate.