1000DF00645 Issue 2
17/26
IVAC® 598 Volumetric Pump
Specifications
Specifications
Electrical Protection
Class I, Type CF.
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
Electrical Safety
IEC/EN60601-1;
Earth Leakage Current (Normal Condition) Maximum 100µA rms (ungrounded) at 220V rms.
Classification
Continuous mode of operation, Portable Equipment.
AC Power Supply -
220 to 240V AC; 50/60 Hz; 0.07A; 3-wire grounded system.
Fuse Type -
2 X T63 mA 250V.
Dimensions -
133mm (w) x 193mm (h) x 184mm (d). Weight: approx. 2.7kg (excluding power cable).
Protection against fluid ingress -
IPX1 - Protected against vertically falling drops of water.
Environmental Specifications
Condition
Operating
Transport and Storage
Temperature
+10°C to +40°C
-40°C to +70°C
Humidity
30% to 75% non-condensing
0% to 100%
Atmospheric Pressure
700hPa - 1060 hPa
500hPa - 1060hPa
System Accuracy:
±5% under specified conditions.
Note:
Accuracy specified under the following conditions:
Head height: 30"/76.2cm.
Test solution: distilled water.
Environmental: ambient temperature.
Back pressure: 20 gauge needle.
Infusion Set: Model 59073 (59073 has been superseded by G59073, which is identical to the previous infusion set).
A
Use of high viscosity liquids or non CareFusion branded sets may adversely affect accuracy
Starting the Infusion / Set-up
Infusion Parameter
Range
Infusion Rate
1 - 999ml/h in steps of 1ml/h
VTBI
(0 - OFF), 1 - 9999ml in 1ml increments
VI (Total)
0 - 9999ml
KVO rate
5 ml/h or current set rate if less than 5 ml/h in primary operation
Battery Specifications
Rechargeable Sealed Lead Acid. Automatically charges when the pump is connected to AC power.
Battery Life - 6 hours @ 125ml/h.
Battery Charging - 24 hours.
Alarm Conditions
Alarm conditions cause the pump to display specific alarm messages, sound an audible alarm, and except for the low battery and KVO
mode alerts, cease operation. Alarm and display messages include: FLO, door, hold, bat., batt., OCCL, RELOAD, Air and Err.
(See the 'Alarms and Display Messages' section of this DFU for message descriptions.).