Delta Speed (km/h): Speed increment pushing “+ speed” key;
Delta Inclination (%): Inclination increment pushing “+ inclination” key;
Autostop pause (sec): reset time;
Warm up time (mm:ss): warm up time;
Cooldown time (mm:ss): cooldown time;
Warm up speed (km/h): warm up speed;
Cooldown speed (km/h): cooldown speed;
Enable WiFi conection: This feature is not available;
Enable RS232 connection: Enable/Un-enable the RS232 connection by cable (it’s necessary to switch off and on
the machine to memorize the new set up);
Enable reverse speed: Enable/Un-enable the reverse speed;
Enable Beep: Enable/Un-enable beep sound at key pressure;
Beep over Cardio HB: Enable/Un-enable beep sound alarm when HB level goes over the HB pre-set max value;
Max reverse speed (km/h): max reverse speed value.
Use the arrow to return back or push CONFIRM key to confirm the selection.
Default per Test:
It’s possible to see or set up some default values as :
CHR initial speed (Km/h): initial speed of CHR test;
CHR duration (sec): time of CHR test;
CHR Freq. add speed (sec): updating time of speed during CHR test;
CHR Incr. speed step: number of time that “Delta speed” is changed during CHR test;
CWL Duration (sec): time of CWL test.
Use the arrow to return back or push CONFIRM key to confirm the selection.
Input protocol:
It’s possible to see or set up the protocol name of communication by RS232 output, the serial speed, the serial delay
(ms) and the unit of measurement speed (kmh/mph).