background image

21

 

If the unit is left idle for a period of approximately 30 seconds during service mode, the unit will exit service mode and return to  

 

normal operation.

 Note

 If an alarm occurs while in the service mode, it will not show until exiting this mode.

 

To reset values back to factory settings, perform the following procedure:

 

 

1. Turn keyswitch to “OFF” position.

 

 

2. Hold the up (         ) and down (         )  arrows while turning the keyswitch to the “ON” position.

 

 

3. Continue to hold buttons for  approximately 10 seconds.

 

 

4. Release buttons and turn keyswitch to  “OFF” position.

 

 

5. Turn keyswitch to "ON" position.

 

 

Performing this reset will overwrite any changes made after receipt of this unit and may overwrite any changes that were  

 

 

made at the factory to optimize performance.

9.5 Temperature Settings

 

The factory default temperature setting is -30°C for plasma freezers. To change the factory temperature settings, refer to the    

 

instructions in 

Section 9.3

.

 

 

The freezer described in this manual is designed for optimum performance at -30°C. It is advisable to call 

 

 

Technical Service before changing setpoints.

9.6 Alarms

 

The alarm system is designed to provide visual and audible warning signals for both power failure and rise in temperature.  

 

The alarm is equipped with a battery backup.

 

Default cold and warm alarm values are -35°C and -20°C. These values may be adjusted, following instructions in 

Section 9.3

.

 

The alarm system is activated only when the key switch is turned to the Alarm On position. The audible warning signal sounds  

 

when there is a power failure, temperature alarm condition, or when the door is ajar for approximately 3 minutes with default door  

 

alarm setting.

 

The Mute function (pressing the ( 

SCAN 

 ) button) allows you to turn off the audio warning without turning off the visual  

 

indicators. The alarm will ring back after approximately 6 minutes if the alarm is still active.

 

During a warm alarm, a flashing of the uppermost bar of the thermometer, the temperature display, and the speaker icon will 

 

occur at the same time. If there has been a warm alarm since the last alarm reset, but the temperature is not currently in an  

 

alarm state, the uppermost bar of the thermometer and the speaker icon will slowly alternate in flashing.

 

During a cold alarm, a flashing of the bulb of the thermometer, the temperature display, and the speaker icon will occur at the  

 

same time. If there has been a cold alarm since the last alarm reset, but the temperature is not currently in an alarm state, the  

 

bulb of the thermometer and the speaker icon will slowly alternate in flashing.

 

During a power failure, the power failure icon will illuminate, the thermometer will display without the bulb, and the  

 

temperature will flash approximately every 3 seconds. If there has been a power failure since the last alarm reset, the power  

 

failure icon and the speaker icon will slowly alternate in flashing.

 

During a low battery condition, the low battery icon will illuminate.

 

The alarms can be reset by pressing and holding the          and           simultaneously.

Summary of Contents for CAX1230LA

Page 1: ...Cardinal Health Premier Plus 30 C Plasma Freezers Installation and Operation CAX1230LA CAX1230LD CAX2330LA CAX2330LD CAX3030LA CAX3030LD CAX5030LA CAX5030LD...

Page 2: ...ubject to change without notice Cardinal Health makes no representations or warranties with respect to this manual In no event shall Cardinal Health be held liable for any damages direct or incidental...

Page 3: ...Specifications 7 7 Installation 8 Location 8 Wiring 8 Shelves Optional 9 Drawers 9 Baskets 12 Temperature Sensor Bottles 12 Door Operation 13 Remote Alarm Optional 13 Final Checks 14 8 Startup 15 Ini...

Page 4: ...Optional 24 Powered Temperature Transmitter 24 Unpowered Temperature Transmitter 24 12 Maintenance 25 Cleaning the Cabinet Interior 25 Cleaning the Condenser Filter 25 Cleaning the Condenser 25 Automa...

Page 5: ...nd installation manual by model number Refer to Section 6 1 for details on the voltage specifications for the various models Table 1 Applicable Models Unit Model Cardinal Health CH1230L A D Cardinal H...

Page 6: ...o moderate injury or damage to the equipment CAUTION This indicates a situation which may result in property damage This symbol indicates surfaces which may become hot during use and may cause a burn...

Page 7: ...isconnect unit from all power sources before cleaning troubleshooting or performing other maintenance on the product or its controls WARNING Risk of Fire This unit is not for storage of flammable mate...

Page 8: ...001 FCC This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection...

Page 9: ...rrier This must be done within 15 days after delivery Then file a claim with the carrier Do not return goods to the manufacturer without written authorization 4 Packing List Inside the freezer cabinet...

Page 10: ...ng areas Professional and clinical blood bank laboratories Pharma and Biotech facilities Industrial and Government facilities or those trained in laboratory protocols put in place at your facility The...

Page 11: ...l Conditions The freezers described in this manual are classified for use as stationary equipment in a Pollution Degree 2 and Overvoltage Category II environment These units are designed to operate un...

Page 12: ...t CAUTION Connect the equipment to the correct power source Incorrect voltage can result in severe damage to the equipment CAUTION Risk of Shock For personal safety and trouble free operation this uni...

Page 13: ...higher and lower positions in the cabinet You can position these slides in the vertical slots which are spaced at one inch intervals The drawers must be removed in order to adjust the position of the...

Page 14: ...from the inside to insert the drawer tabs completely into the slots Make sure both the drawer tabs are aligned with the slots on the slides before pushing the drawer down between the slides CAUTION B...

Page 15: ...of the cabinet 6 Determine desired location for the slide and insert the slide into the rear pilaster at a 45 degree angle towards the center of the cabinet 7 Once the slide is seated in the rear pil...

Page 16: ...talling and changing position of the baskets can be found in Section 7 4 1 Section 7 4 2 and Section 7 4 3 The maximum basket capacity is 27 kg 60 lbs 7 6 Temperature Sensor Bottles The temperature di...

Page 17: ...s of the cabinet to help facilitate routing of tubing independent temperature sensors and other non electrical items The ports are not intended to supply power to devices inside the device Be sure to...

Page 18: ...COMMON and CLOSE ON FAIL wires must be tied together in this application 3 Plug the alarm system service cord into an electrical outlet The contacts will trip in the event of a power outage high tempe...

Page 19: ...to avoid potential frost bite 8 2 Product Loading and Unloading Guidelines When loading your freezer take care to observe the following guidelines Never load the plasma freezer beyond capacity The num...

Page 20: ...perature in degrees Celsius as measured by the primary sensor inside the cabinet You can use the buttons to display other values such as setpoints and highest and lowest recorded temperatures The numb...

Page 21: ...rms are on and the setpoints cannot be modified 13 Power On when the key switch is in the on position the unit power is on with no alarms active 14 Off when the key switch is in the off position the u...

Page 22: ...SEr to indicate that you have entered the service mode followed by the software checksum values flashing on the screen The service icon will also illuminate Use up and down buttons to modify the valu...

Page 23: ...mode This offset is also used on some freezer models as a response todoor openings 5 Press Scan SCAN Cnt oFSt Control Probe Offset The difference in temperature between the control probe value and th...

Page 24: ...mperature drip pan probe can reach before stopping the defrost drip time 16 Press Scan SCAN qUA Quality Mode Normally OFF 00 Used for production line testing 17 Press Scan SCAN bot BOT mode Normally O...

Page 25: ...vated only when the key switch is turned to the Alarm On position The audible warning signal sounds when there is a power failure temperature alarm condition or when the door is ajar for approximately...

Page 26: ...to access the recorder 2 Connect the nine volt DC battery located at the recorder s upper right corner This battery provides back up power 3 Install clean chart paper refer to Section 10 2 at the next...

Page 27: ...ng If AC power fails the LED indicator on the recorder flashes to alert you to a power failure The recorder continues sensing cabinet temperature and the chart continues turning for approximately 24 h...

Page 28: ...Temperature Transmitter Label Description 1 Terminal Strip 2 Rear of Unit 3 Controller and or Indicator 4 Customer Provided System Note The building management system load wiring must have a resistanc...

Page 29: ...ds through the slot in the back cage 2 Shake the filter to remove loose dust 3 Rinse the filter in clean water 4 Shake the excess water from the filter and let it dry 5 Reinstall filter by pushing dow...

Page 30: ...or frost which forms at the point of gasket failure Make sure that the cabinet is level refer to Section 7 1 1 for leveling information Keep the door gaskets clean and frost free by wiping gently wit...

Page 31: ...ect for your unit refer to Table 2 If the outlet is dead check the circuit breaker or fuses The unit should not be connected to a GFCI Ground Fault Circuit Interrupter protected outlet as it may be su...

Page 32: ...to the unit call an electrician Setpoints need to be adjusted To adjust the setpoint refer to Section 9 3 Extended air leak into freezer Warm air entering the freezer from extended door openings leak...

Page 33: ...lowing 1 Remove items and defrost unit Be sure to clean up any biological safety hazards 2 Remove the cabinet door to help prevent entrapment inside of a unit 3 Have a certified technician remove the...

Page 34: ...d without further force or effect if there is any i repair made to the product by a party other than the manufacturer or its duly authorized service representative ii misuse including use inconsistent...

Page 35: ...This page intentionally left blank...

Page 36: ...nal Health Waukegan IL 60085 USA 2021 03 cardinalhealth com 2021 Cardinal Health All Rights Reserved CARDINAL HEALTH and the Cardinal Health LOGO are trademarks of Cardinal Health and may be registere...

Reviews: