12
8. Replace the safety clip by hooking the clip under the front pilaster slide bracket. Then rotate the wire safety clip to position
the other bend of the wire safety clip over the top of the same bracket.
Figure 7. Clip Replacement
Drawer slides do
not
require lubrication. Additional lubricant could impede movement of the drawers when lubricant is cold.
7.5 Baskets
Baskets can be ordered as an accessory for this unit.
The basket slides are adjustable to higher and lower positions in the cabinet. You can position these slides in the vertical slots which
are spaced at one-inch intervals. The baskets must be removed in order to adjust the position of the slides. The procedure for removing,
reinstalling and changing position of the baskets can be found in
Section 7.4.1
,
Section 7.4.2
, and
Section 7.4.3
.
The maximum basket capacity is 27 kg (60 lbs).
7.6 Temperature Sensor Bottles
The temperature displayed on the control panel and chart recorder is measured by a probe inserted in sensor bottle inside the cabinet.
The sensor probe must be inserted in the glycerin bottle prior to operation.
The sensor bottle is positioned on the left wall towards the top of the unit.
To install the sensor(s) into the bottles, complete the following steps:
1. Remove the bottle from the bracket and verify that each bottle is full of liquid.
2. Remove the solid cap from the bottle and save for the next time the unit needs to be moved.
3. Remove the taped lid and temperature sensor(s) from the wall.
4. Insert the temperature probe(s) into the bottle.
5. Secure the lid to the bottle and return the bottle to the bracket.