Electromagnetic Compatibility (EMC)
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The ABPM devices are intended for use in the electromagnetic environment specified below. It is the responsibility
of the customer or user to ensure that the ABPM devices are used in such an environment.
Immunity Test
EN/IEC 60601 Test Level
Compliance
Level
Electromagnetic environment—
Guidance
Electrostatic discharge
(ESD) to EN 61000-4-2/
IEC 61000-4-2
±8.0 kV contact
±2.0 kV air
±4.0 kV air
±8.0 kV air
±15.0 kV air
±8.0 kV
±2.0 kV
±4.0 kV
±8.0 kV
±15.0 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst to EN 61000-4-4/
IEC 61000-4-4
±2.0 kV for power supply
lines
±1.0 kV for input/output
lines
not applicable
not applicable
Main’s power should be that of a
typical commercial or hospital
environment.
Surge to EN 61000-4-5/
IEC 61000-4-5
±0.5 kV differential mode
±1.0 kV differential mode
±0.5 kV common mode
±1.0 kV common mode
±2.0 kV common mode
not applicable
not applicable
Main’s power should be that of a
typical commercial or hospital
environment.
Voltage dips, short inter-
ruptions and voltage
variations on power
supply input lines to EN
61000-4-11/
IEC61000-4-11
0 % power supply
for 10 ms (0.5 cycles)
0 % power supply
for 20 ms (1,0 cycles)
70 % power supply
for 500 ms (25 cycles)
0 % power supply
for 5000 ms (250 cycles)
not applicable
not applicable
not applicable
not applicable
Main’s power should be that of a
typical commercial or hospital
environment. If the user of the ABPM
devices requires continued operation
during power mains interruptions, it is
recommended that the ABPM devices
are powered from an uninterruptible
power supply or a battery.
Power frequency (50/
60 Hz) magnetic field to
EN 61000-4-8/IEC
61000-4-8
30.0 A/m
30.0 A/m
Power frequency magnetic fields
should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.