Fehlermeldungen
5
Error Codes
E 03
Internal hardware error. Please contact your
local authorized dealer
(http://www.bparo.com).
E 04
Batteries depleted. Code appears when the
battery capacity is insufficient for new BP
measurements.
E 05
Measurement time over. Code is displayed
after measurement duration of 180 seconds
(without inflation time).
E 06
This code appears when
−
- the current cuff pressure would exceed the
maximum permissible inflating pressure of
280 mmHg,
−
- the device has reached the selected
maximum inflating pressure,
- a measurement is not possible (device waits
until the next time of measurement).
E 07
Inflation time over. The maximum inflation
time of 130 seconds has elapsed. This
condition indicates a leak in the cuff or
tubing, or a defect at the cuff connector.
E 08
Insufficient number of oscillations detected:
For a correct measurement, the system must
detect a minimum of 8 oscillations.
For
deflation measurement method:
Tighten the cuff so that one finger, but not
two, can be inserted between the patient's arm
and the cuff. At the same time the device
switches to a deflation rate of 4 mmHg/s.
When it detects more than 13 oscillations later
on, the rate changes to 6 mmHg/s.
For
inflation measurement method
:
This error message will not be displayed
because the ABPM devices automatically
switch to the deflation measurement method if
the number of detected oscillations is
insufficient.
E 10
Memory space is full. 400 pressure
measurements have been taken and thus the
storage capacity is exhausted.
E 11
Motion artifact during diastole detection.
E12
Diastolic reading outside measuring range.
E 20
Systolic reading outside measuring range.
(Codes E12 and E20 are displayed when the
systolic and diastolic values are outside the
range in which oscillations were detected.)
E 21
Systolic reading below measuring range.
E 22
Systolic reading above measuring range.
E 23
Motion artifact during systole detection,
air discharge speed too high, e. g. due to
leakage.
E 24
Difference between systolic and diastolic
pressure too small (10 mmHg or less).