26
BPAro
ENG Revision A
Order Information
7
Technical Specifications
Measuring Method
−
Oscillometric, selectable measurement method:
deflation measurement method or inflation
measurement method
Measuring Interval
−
2 - 120 min, programmable
Maximum cuff pressure
−
300 mmHg
Interfaces
−
USB (HID)
Dimensions
−
Height: 27 mm
−
Width: 80 mm
−
Depth: 105 mm
Battery
−
2 Mignon (size AA) rechargeable NiMH batteries,
1,2 V, > 1.500 mAh or
−
2 Mignon (size AA) alkaline batteries
Battery Charge Time
−
2 - 3 hours
Battery Charger
−
protection class II, IP20
−
100 - 240 VAC 50/60 Hz, 0.5 A
Ambient Conditions
Operation
−
tempe 5 °C to + 40 °C
−
relative humidity 15 - 93 %, no condensation
−
atmospheric pressure between 700 - 1060 hPa
−
altitude (relative to sea level) -400 - 2800 m
Transport and Storage
−
temperature -25 to +70 °C
−
relative humidity 10 - 93 %, no condensation
−
atmospheric pressure 500 - 1060 hPa
−
altitude (relative to sea level) -400 - 4500 m
Protection Class
−
IP20: ABPM device
−
IP02: wearable pouch of the ABPM device
−
IP22: ABPM device in wearable pouch
Expected Service Life
−
ABPM device: 10 years
−
Cuff: 20.000 cycles of reapplication
7.1
Blood Pressure Measurement
Measuring Range
−
systolic pressure
60 - 260 mmHg
−
diastolic pressure
40 - 220 mmHg
−
mean pressure
45 - 250 mmHg
−
pulse rate (HR)
35 - 240 min
-1
Measurement Accuracy (determined in a clinical
study according to ISO 81060-2)
−
deflation measurement method:
systole:
0,7 ± 2,5 mmHg
diastole:
0,5 ± 2,2 mmHg
pulse rate (HR):
1,0 ± 2.6 min
-1
−
inflation measurement method:
systole:
0,9 ± 3,8 mmHg
diastole:
0,6 ± 3,2 mmHg
pulse rate (HR):
0,6 ± 3,5 min
-1
Measurement Capacity
−
up to 400 blood pressure measurements
Weight
−
< 225 g, incl. batteries
Note
The device needs 30 min to get ready for its
intended use and reach the operation
conditions form the minimum storage
temperature and the maximum storage
temperature, if the room temperature is 20 °C.