Cleaning, Maintenance, Disposal
BPAro
ENG Revision A
6
Cleaning, Maintenance, Disposal
6.1
Cleaning and Disinfection of the
Equipment Surface
•
Switch off the ABPM device.
•
Wipe the device down with a soft, lint-free cloth for
cleaning. Liquids must not penetrate the device.
Spray disinfection has proved successful. Incidin®
Foam or equivalent disinfectants that are used in
practices or hospitals are suitable (Please respect
the information of the manufacturer especially
regarding the exposure time).
6.2
Cleaning and Disinfection of the Cuffs
−
Light staining can be wiped off with a damp cloth.
−
In case of heavy staining, wash the cuff with soapy
water or disinfectant detergent (not in the washing
machine). No liquid may penetrate into the cuff
bladder or the connecting hose.
−
After the application, spray disinfection has proved
successful.
Incidin®
Foam
or
equivalent
disinfectants that are used in practices or hospitals
are suitable (Please respect the information of the
manufacturer especially regarding the exposure
time).
−
After cleaning, rinse the cuff thoroughly with clean
water and let it dry in the air at room temperature
for about 15 hours.
6.3
Cleaning and Disinfection of Cables
•
Disconnect cables from the device before
cleaning.
•
Use a cloth moistened with soapy water to wipe the
cables clean. Do not immerse cables in liquid.
Warning
Shock Hazard —
Disconnect the ABPM device from the PC or
printer before cleaning.
Caution
Equipment Damage —
Do not disinfect the device surface with phenol-
based disinfectants or peroxide compounds.
Warning
Shock Hazard, Equipment Damage —
Equipment into which liquids have entered must be
inspected by a service technician before use.
werden.
Warning
Risk to Persons —
Equipment and accessory have to be disinfected
between the uses on different patients. Additionally
national regulations for the cleaning and
disinfection have to be considered