![Capintec CRC-712MH Owner'S Manual Download Page 44](http://html1.mh-extra.com/html/capintec/crc-712mh/crc-712mh_owners-manual_494183044.webp)
CAPINTEC, INC.
CRC
®
-712MH
TROUBLESHOOTING
Some problems may be very easy to diagnose and correct in the field with little or no
equipment. If a problem should occur, check here before you call for service. You may be
able to save a considerable amount of time and money.
Nothing appears on the display
•
Make sure calibrator is plugged into a live outlet and is turned on.
•
Check fuse and replace if necessary. See: LINE VOLTAGE SELECTION/FUSE
High background indication
•
Chamber Well, liner, or dipper may have become contaminated. See: CHAPTER
5: QUALITY ASSURANCE & ACCEPTANCE TESTING, SECTION:
CONTAMINATION TEST.
•
Background may actually be high. Check by removing the dipper and placing a
lead sheet over the top of the well.
Indication of significant negative activity
•
Zero setting or Background level may have changed. Perform the Daily Test. See:
CHAPTER 5: QUALITY ASSURANCE & ACCEPTANCE TESTING.
Printer will not respond
•
Make sure printer is plugged into a live outlet, turned on, and “selected”.
•
Make sure that either paper or a ticket is in the paper path.
7 - 4
CLEANING AND MAINTENANCE