CAPINTEC, INC.
CRC
®
-712MH
GENERAL OPERATIONAL SETUP
Two configuration changes may be preferred or required for normal operation of the CRC
®
-
712MH.
•
Locking
the Units of measurement in the Curie or Becquerel mode, and
•
Reassignment
of one or more of the preset isotope selection push-buttons.
Locking the Curie or Becquerel Mode of Operation
The CRC
®
-712MH provides radioisotope measurements in Curies (Ci) or Becquerels (Bq) as
the selected units of measurement by the
CURIE / BECQUEREL
selector switch on the front
panel of the instrument. This feature is particularly useful when a country is changing from
the older unit of measurement the (Curies) to the currently recognized unit of measurement
the (Becquerels). When this transition is complete, it may be desirable to permanently lock
the selector switch in the Becquerel mode.
1. Facing the rear of the calibrator, locate the Unit’s Selector Switch Locking Screw and
remove it from its storage position (Refer to Figure 3-2).
2. Now moving to the front panel, rotate the Selector Switch to the desired unit of
measurement.
3. Insert the locking screw and tighten. (Refer to Figure 4-2)
Figure 4-2
January 08