PREVENTATIVE MAINTENANCE
The Quality Assurance Tests described in CHAPTER 5: QUALITY ASSURANCE &
ACCEPTANCE TESTING should be performed periodically as stated.
Tests must be performed in an environment where the temperature is stable within a range
of +50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing. The unit should be powered-up for at least one-half hour prior to performing any
measurements. No other precautions need to be observed.
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
If the unit fails to pass any of the tests, the user
should not
attempt to perform any
adjustments to the system. In this event, please contact Capintec for further assistance.
The Quality Assurance tests should be immediately performed if:
•
The equipment has been subjected to extreme physical stress,
•
Liquids enter the readout unit, and/or chamber, or
•
Any cable shows signs of damage
SERVICING
The system is covered by a one year limited warranty, under normal conditions of use.
Other than the Readout Unit fuses (see LINE VOLTAGE SELECTION/FUSE
REPLACEMENT in this chapter), there are no user serviceable parts contained in the
system.
Every five years, the system should be returned to Capintec’s Authorized Service Center for
a complete verification.