Validation
Some industries require sampling equipment to be validated before each
use. For the 100 L/min MicroBio MB2 and MB2-RSH this can be achieved
using the MicroBio Validation Kit, part number A-00058 available from
Cantium Scientific Limited or your local distributor. For the MB2-HiFlow,
validation can be achieved using the Qualisair™ qCR kit, Cantium Scientific
Limited part number A-00422.
Calibration
It is recommended all models of MicroBio MB2 are regularly calibrated in
accordance with specific industry best practice. Typically, this will be on the
anniversary of the instrument entering service.
The only calibration adjustment on the instrument is the flow rate.
Calibration is normally performed at an air pressure of 1013 mbar at 20ºC.
For calibration services please contact Cantium Scientific Limited or your
local distributor. For in-house calibration, we recommend the use of the
Qualisair™ qCR kit.