Inserting a Contact Plate or Petri Dish
Carefully remove the sampling head cover from
the MicroBio MB2, unless it is being carried
separately in a sterile container. Only hold the
edge of the head.
Do not touch the perforated
or inside surfaces.
With the sampling head removed, insert a petri
dish/contact plate inside the springs, so that its
base sits firmly on the support plate, then re-fit
the sampling head.
The standard MicroBio MB2 can use 220 or 400
hole sampling heads. The 220 hole sampling head
can be used with both 55mm/65mm contact plates
or 90mm petri dishes, but only the centre 55mm of
the petri dish will collect samples. Using the 400 x
0.7mm hole sampling head will give nearly full
coverage of a 90mm petri dish, improving reliability
of counts.
The MicroBio MB2-HiFlow model can only use
90mm Petri dishes and the 400 x 1mm hole
sampling head.
ALWAYS STERILISE THE SAMPLING HEAD
PRIOR TO EACH USE
14