Guidance and manufacturer’s declaration for EMD
54
The AX-B2735W is intended for use in the electromagnetic environment specified below.
The customer or the user of the AX-B2735W should assure that it is used in such an
environment.
Table 2
Electromagnetic immunity
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment – Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±(2, 4, 8, 15) kV
air
±8 kV contact
±(2, 4, 8, 15) kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be
that of a typical commercial
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be
that of a typical commercial
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
0% U
T
; 0.5 cycle
0°, 45°, 90°,
135°, 180°,
225°, 270°, and
315°
0% U
T
; 1 cycle
and 70% U
T
; 25/
30 cycles
Single phase: at
0°
0% U
T
; 250/300
cycles
0% U
T
; 0.5 cycle
0°, 45°, 90°,
135°, 180°,
225°, 270°, and
315°
0% U
T
; 1 cycle
and 70% U
T
; 25/
30 cycles
Single phase: at
0°
0% U
T
; 250/300
cycles
Mains power quality should be
that of a typical commercial
environment. If the user of the
AX-B2735W requires
continued operation during
power mains interruptions, it is
recommended that the AX-
B2735W be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
000_DIA-2019-002-E04_AX-B2735W.book Page 54 Friday, November 27, 2020 12:22 PM
Summary of Contents for AX-B2735W
Page 6: ...Conventions 6 000_DIA 2019 002 E04_AX B2735W book Page 6 Friday November 27 2020 12 22 PM ...
Page 21: ...2 Introduction 000_DIA 2019 002 E04_AX B2735W book Page 21 Friday November 27 2020 12 22 PM ...
Page 41: ...6 Maintenance 000_DIA 2019 002 E04_AX B2735W book Page 41 Friday November 27 2020 12 22 PM ...
Page 43: ...7 Specifications 000_DIA 2019 002 E04_AX B2735W book Page 43 Friday November 27 2020 12 22 PM ...
Page 59: ...000_DIA 2019 002 E04_AX B2735W book Page 59 Friday November 27 2020 12 22 PM ...