The profile may not be completed. Verify that the Remaining Time
for the profile is 0:00.
The instrument may not be communicating with the computer. Log
into VISCPRO as a Manager and click Service/MRV/Monitor to
open the MRV monitor. If the data window is updated regularly with
new data, then communication is being maintained.