Proceed as follows:
1.
Remove the applicator cover by pulling it off. Discard the used cover and filter assembly.
2.
If it becomes necessary to clean the applicator cover during treatment, use hot tap water
(no more than 122°F/50°C). The cover should be flushed and washed thoroughly in hot
water and mild disinfectant soap or detergent that do not contain alcohol, until all residual
VelaSpray Ease is removed. Wipe and dry the cover with a soft cloth. Reconnect the
cover only when completely dry.
3.
Clean the Vcontour applicator's components at the end of every day as follows:
•
Refer to Figure 26: clean the RF applicator contacts (A), the ID microchip contact
points (B) and the vacuum inlet (C) with a clean, soft cloth.
•
Gently clean the IR lamp's aperture (D) using a gentle optical lens cleaning
Figure 25: Replacing the Vsmooth Applicator's Filters
∆
Caution
Make sure that the interior of the applicator is completely dry before
connecting a new replaceable cover.