PVS5120-Series Portable Samplers
33
FIGURE 8-6. SDI-12 Sampler Settings screen
2. Click the
Advanced Sampler Settings
tab (FIGURE
). In most cases,
only the Enable Pressure Relief setting is changed; reasons to change the
other options from factory defaults are discussed in Section
(p. 13)
.
a. Set
Sample Chamber – Pressure Relief
. If collecting a sample
volume >450 mL or using an intake hose longer than 50 ft, this feature
must be enabled.
3. Click
Apply
. It is a good idea to save the configuration file for future
reference.
4. Connect the PVS5120CBL cable to the front of the sampler, and wire the
cable to the datalogger (TABLE
5. Use
LoggerNet
to connect to the datalogger. The following steps assume a
knowledge of how to setup the datalogger and connect either via a direct
connection or remote connection.
a. From the
LoggerNet Connect
window, click
Connect | Collect Now
to collect any data on the datalogger.
b. From the
Connect
screen, click the
Send New
button. Browse to find
the
CRBasic
program that controls the sampler via SDI-12 (see
. Click
Open | Send
.