e. Typically leave the
Never Full Operation
disabled. The
Never Full
Operation
is typically enabled only when the sampler is easily
accessible and containers are collected regularly, for instance at a
waste water treatment facility.
f. If the Auto Deploy routine has been run, the
Purge Time
is already
set. If not using Auto Deploy, enter the amount of time the sampler
hose is evacuated before and after the sample is collected. The
maximum value is 800 s. The final purge time can be set separately
(p. 13)
g. If a
Water Present Sensor
is present, click to install. This sensor is
wired to the sampler control cable. See Section
, for more information.
FIGURE 8-11. PakBus® Setting Editor screen
6. Click the
Advanced Sampler Settings
tab (FIGURE
). In most cases,
the only setting a user may want to change is to Enable Pressure Relief
setting.
a. Set
Sample Chamber – Pressure Relief
. If collecting a sample
volume >450 mL or using an intake hose longer than 50 ft, this feature
must be enabled.
7. Click
Apply
. It is a good idea to save the configuration file for future
reference.