SDI-12 Measurements
The CS650 responds to SDI-12 commands
M!
,
M1!
,
M2!
,
M3!
,
?!
, and
I!
shows the values returned for each of these commands.
, for SDI-12 wiring details.
This section briefly describes using the SDI-12 commands.
Additional SDI-12 information is available at Appendix
(p.
or
www.youtube.com/user/CampbellScientific
Table 8-4. CS650 SDI-12 Commands
SDI-12 command
(“a” is the sensor
address)
Values Returned
Units
aM!
1)
Volumetric Water Content,
2)
Electrical Conductivity,
3)
Temperature
m
3
/m
3
dS/m
°C
aM1!
1)
Permittivity,
2)
Electrical Conductivity,
3)
Temperature
dS/m
°C
aM2!
1)
Period,
2)
Voltage Ratio,
3)
Temperature
Sec
°C
aM3!
1)
Volumetric Water Content,
2)
Electrical Conductivity,
3)
Temperature
4)
Permittivity,
5)
Period,
6)
Voltage Ratio,
m
3
/m
3
dS/m
°C
Sec
aM4! .. aM9!
No Values Returned
?!
Returns the SDI-12 Address
aI!
CampbellSci, OS version, Product
Serial Number
Up to 10 CS650 probes may be connected to the same datalogger control port as
long as each one has a unique SDI-12 address. The CS650 ships with a default
SDI-12 address of 0 unless otherwise specified at the time of ordering. The SDI-
12 address may be changed through
DevConfig
and Device Configuration Utility
) or with a terminal emulator in SDI-12
transparent mode (see Appendix
).
SDI-12 communication is established using the
SDI12Recorder
() CRBasic
instruction. See Appendix
, for more detail on
SDI-12 communication.
8.2.1
Use of Multiplexers
Multiplexers such as Campbell Scientific’s AM16/32B may be used to connect up
to 32 CS650 probes to a single control port. When using multiplexers, the
simplest configuration is for all probes to have the same SDI-12 address.
NOTE