![background image](http://html.mh-extra.com/html/campbell/cs451/cs451_instruction-manual_489463019.webp)
CS451/CS456 Submersible Pressure Transducer
Transparent mode is entered while the PC is in telecommunications with the
datalogger through a terminal emulator program. It is most easily accessed
through Campbell Scientific datalogger support software, but is also accessible
with terminal emulator program.
To enter the SDI-12 transparent mode, enter Terminal Emulator from
LoggerNet, PC400 or PC200W datalogger support software. A terminal
emulator screen is displayed. Click the
Open Terminal
button.
For CR800-series, CR1000, CR3000 dataloggers, press <Enter> until the
datalogger responds with the prompt (“CR800>” for the CR800). Type
SDI12
at the prompt and press <Enter>. In response, the query
Enter Cx Port 1,3,5
or 7
will appear. Enter the control port integer to which the transducer is
connected. An
Entering SDI12 Terminal
response indicates that SDI-12
Transparent Mode is active.
FIGURE 7-3. Terminal Emulator
For CR10X and CR510 standard mixed-array dataloggers, the
telecommunications command to enter SDI-12 transparent mode is
nX<Enter>
where
n
is the control port being used for SDI-12. In this example, the selected
control port is
C1
, so the command would be
1X<Enter>
. In response, the
datalogger opens the link to control port 1 and responds with a prompt. CR10X
and CR510 dataloggers reply with
entering SDI-12
.
CR10X-PB and CR510-PB dataloggers require
*#n
being entered at the
prompt, where
n
is the control port being used.