CS451/CS456 Submersible Pressure Transducer
7.1.2 Device Configuration Utility (version 2.03 or higher)
The Device Configuration Utility allows you to change the settings of the
CS451/CS456. Device Configuration Utility is shipped on the Campbell
Scientific ResourceDVD included with the CS451/CS456.
To use Device Configuration Utility, the transducer needs to be connected to
the PC via the A200 (see Section 7.1.1,
PC Connection Using the A200
). After
installing Device Configuration Utility and connecting the transducer to the
PC, select CS451 from the Device Type list on the left column of the screen.
In the PC Serial Port box, select the COM port that was assigned to the A200
(see Section 7.1.1.4,
Determining which COM Port the A200 has been
Assigned
). Click on the
Connect
button to enable communication with the
sensor. Once successfully connected, the screen should look like FIGURE 7-2.
FIGURE 7-2. Connect screen
There are three settings that can be changed: SDI-12 address, Pressure/Level
Units and Temperature Units. Double-click on the window of the units to be
changed. This will open a pick menu box. Select the desired units and
Apply
the changes.
7.1.3 SDI-12 Transparent Mode
Transparent Mode allows direct communication with the CS451/CS456. This
may require waiting for programmed datalogger commands to finish before
sending responses. While in the transparent mode, datalogger programs may
not execute. Datalogger security may need to be unlocked before transparent
mode can be activated.