53
Select option 1 to start the calibration. You will then be asked to confirm that you
would like to perform a calibration. Please note, once you have entered yes at this
point you will not be able to exit until the test is complete. However, power
cycling the unit at this point will have no adverse effect on the sensor.
At this stage it is advisable to clean the lenses. Refer to Section 18
`Cleaning’ for more information. A simple visual check may be
enough to confirm the lenses are clean.
Once you have started the tests you will be asked for the sensor calibrator serial
number and extinction coefficient with a confirmation at each step giving you the
chance to correct typing mistakes.
When asked for confirmation you do not need to press return after
you type ‘y’.
NOTE
NOTE
Starting calibration.
Input the sensor calibrator serial number ->E2002
Is E2002 correct? (Y/N)?
Input the sensor calibrator constant ->28.8
Is 28.8 correct? (Y/N)?
Place one calibration bung into each hood, then
press any key.
Do you want to perform a calibration Y/N?