Menu 3:
The calibration menu
Consult the ‘calibration’ section of this manual for information on how to
calibrate the sensor.
Once a calibration is finished changes are immediate, but factory
calibrations can be restored if needed using Option `(2)’ in the
calibration menu. The disk constants however remain as the last
disk used.
Option (4) gives the systems information menu containing useful information such
as temperature and system alarms.
The parameters `calibration value factory offset’ and `calibration value factors
scale’ are the factory calibration coefficients from the last factory calibration. The
parameters `calibration value cal offset’ and `calibration value cal scale’ are the
calibration coefficients from the latest calibration.
NOTE
CS125 CALIBRATION - MENU 3
ID 0
S/N 1003
(1) Perform calibration
(2) Restore the factory calibration
(3) Perform dirty windows zero offset calibration
(4) Restore dirty windows factory calibration
(9) Refresh
(0) Return to main menu
->