A4 - Test Watch
These functions are not required in normal use of the
MotionWatch. Do not use unless instructed to do so by a
CamNtech representative.
A5 - Changing the
Battery
The MotionWatch has been designed to facilitate quick
and simple battery changing.
The change the battery, open the MW8 casing as
described in Section 1.1 An Overview of the
MotionWatch System to expose the USB
connection and battery.
Do not connect the USB cable!
Using a suitable tool, carefully push the battery out
from the holder towards the USB connector.
Grasp the battery and withdraw it completely from
the holder.
Wait at least 1 minute
or press the marker button
3 times
then place a fresh CR2032 battery into the
holder (ensuring correct orientation!) and push
back until fully home.
Note: Data will NOT be lost by removal of the battery.
Observe the LED on the front of the MotionWatch, the
correct start sequence is indicated as follows:
LED stays on for 4 seconds.
LED goes off for 8 to 12 seconds.
LED flashes once.