MC-II Flow Analyzer
19
6.
The upper display will show the prompt
CodE
and the lower display
will show
no
.
Press
ENTER/STEP
to accept the factory default
no
setting, disabling the security code feature.
If
YES
appears in the lower display, press
INCREMENT
until
no
is shown. (See “Enable Password Security Option,” page 11, for more
information on security settings
.
) Press
ENTER/STEP
to accept the
selection.
7.
The prompt
SEt.tot
will appear in the top display, and
no
(default
setting) will appear in the lower display.
Since no preprogrammed total is to be entered for this example, press
ENTER/STEP
to bypass this menu selection.
The prompt
SAVing
will appear in the lower display while the
calibration settings are saved to nonvolatile memory. The MC-II will
automatically return to the Run mode.
In Run mode, press
DEC. POINT
to position the decimal point in the
top (volume) display to show tenths.
Deselect the Password
Security option.
Bypass the Preset Total
option.