MC-II Flow Analyzer
13
barrels
gallons
Recalibrating Your MC-II
If your MC-II was purchased as a system, it was calibrated at the factory
with its companion NuFlo turbine flowmeter. Verify the calibration set-
tings to ensure that no further calibration is required prior to use.
If you purchased a stand-alone MC-II without a flowmeter, the MC-II
must be calibrated before being operated.
The following information is needed to calibrate the MC-II:
•
unit of measure for volume
•
flowmeter calibration factor (in pulses per gallon)
•
unit of measure for flow rate
The calibration procedure is explained in detail below.
1.
Press and hold
ACCESS
for 3 seconds, then release, to enter Cali-
brate mode. Following a quick diagnostic test,
Prog.no
will appear
in the top display and the version number of the firmware will appear
in the bottom display for 1 to 2 seconds.
The
tot.Eng
menu will then appear in the top display. The factory
default volume unit - bbL - will appear in the bottom display.
IMPORTANT: If the pulse output option is in use, you must select
uSEr
as the unit of measure for volume. Any other selection will
cause the pulse output to yield an incorrect number of pulses.
2.
To select a unit of measure for volume: bbl, gal, m
3
, liter, or user-
defined, press
INCREMENT
repeatedly until the appropriate unit
of measure appears in the bottom display. Press
ENTER/STEP
to
accept the selection.
•
If you select a preprogrammed unit, the
Pu.P.gAL
menu will ap-
pear in the top display, allowing you to enter the calibration factor
of your companion flowmeter. The MC-II will automatically
calculate the divisor. Proceed to Step 3 for further instruction.
•
If
uSEr
is selected,
Ent.diV
will appear in the top display,
prompting you to enter a calculated divisor in terms of the unit
specifications. Enter the divisor, from right to left, using the
IN-
CREMENT
key to select each digit, and the
ENTER/STEP
key
to accept each digit selection. Then proceed to Step 4 for further
instruction. An example of MC-II calibration with user-defined
units is provided on page 20.
Press ACCESS to enter
Calibration mode.
TOTAL
RATE
Set unit of measure for
VOLUME.
Factory default setting is bbl.
cubic
meters
liters
user-defined
TOTAL
RATE
Enter the calculated divi-
sor for the user-defined
volume unit.
(Applicable only when
user-defined unit is se-
lected for volume.)
Summary of Contents for NuFlo MC-II
Page 1: ...NUFLO MC II FlowAnalyzer User Manual Manual No 100079666 Rev OE ...
Page 6: ...MC II Flow Analyzer 2 Figure 1 Nomenclature exterior ...
Page 7: ...MC II Flow Analyzer 3 Figure 2 Nomenclature interior ...
Page 32: ...MC II Flow Analyzer 28 ...
Page 38: ...MC II Flow Analyzer 34 ...
Page 44: ...40 MC II Flow Analyzer ...
Page 45: ......
Page 46: ......