PEDICLE PREPARATION & DILATION
Insert the Cannulated Pedicle Probe (CM-129-01) over the K-wire (CM-130-XX). Twist the probe
into the pedicle to open the pathway into the vertebral body. Use the succession of etched
lines on the K-wire (CM-130-XX). The depth markings are etched in 15mm increments. Note the
length of K-wire (CM-130-XX) protrusion through the handle of the Pedicle Probe (CM-129-01) and
ensure that the K-wire (CM-130-XX) does not advance anteriorly during pedicle preparation. The
K-wire (CM-130-XX) Gripper (CM-102-02) can be used to secure the K-wire (CM-130-XX) position.
DILATION OPTIONS:
1.
If using the 1-Step Dilator assembly, slide 1-Step Inner Dilator (CM-119-05) through 1-Step Outer
Dilator (CM-119-06).