Find the Front Side indicator marking on the Rod Inserter. The
rod engages the Rod Inserter from this side only.
The rod will
not lock to the Rod Inserter if loaded backwards
. Once
the rod is fully seated within the Rod Inserter, thread the dial
counterclockwise finger tight. Once finger tight, the rod will be
rigid to the Rod Inserter.
Front Side
Laser etched
Marking
Slowly insert the rod into the most superior pedicle screw (as shown) or most inferior pedicle screw.
Ensure the bulleted tip of the rod is below the fascia tissue layer.