TLC Scanner 4
Instruction Manual, June 2018
Page 17
5.6
Maintenance data sheet
CAMAG Maintenance data sheet
TLC Scanner 3/4
April 2012/UB
Purpose
The maintenance data sheet informs about maintenance interval of the respective instru-
ment as well as the proposal for IQ/OQ interval if applicable. In addition, it identifies con-
sumable parts with the respective replacement cycle.
Maintenance interval
Mainte-
nance
12 Months
IQ/OQ
12 Months
Consumable parts
Part No.
Description
Replacement cycle
027.6440 Hg lamp
Upon need*
027.6441 D2 lamp
Upon need*
027.6442 W lamp
Upon need*
352.0014 Compartment illumination UV254
Upon need
115.0123 L-shaped magnetic strip
Upon need
115.0124 Magnetic strip
Upon need
* Latest when the intensity is dropping under 60%
Summary of Contents for 027.6200
Page 1: ...INSTRUCTION MANUAL TLC SCANNER 4 ...