Step 17: Your system is now ready to be plugged into the appropriate GFCI protected outlet. Plug the unit
in and refer to the “System Disinfection” section before any water is allowed to flow through the system.
System Disinfection:
Now that the system is installed you will need to disinfect the lines throughout the entire home or facility.
To accomplish this, please follow these simple steps:
Step 1: Before performing this task, check to make sure that there are no “dead ends” in the lines within
the water system you are treating as these can harbor bacteria. Also ensure that the UV system is
powered-‐up and ready for operation before you start this disinfection process.
Step 2: Remove the cartridge from the filter sump and fill with 1-‐2 cups of household bleach (most
household bleaches are 5.25% chlorine). Replace filter sump and slowly turn on the water supply.
Step 3: Go to each location in the home or facility (including outdoor taps) and run some water until
chlorine can be detected (by smell). Ensure all faucets, dishwasher, shower heads, washing machine,
toilets, showers, refrigerators, etc. are all checked. Once the chlorine is detected at each location close all
faucets and let the chlorine sit in the lines for a minimum of 30 minutes.
Step 4: After 30 minutes, reinstall the filter cartridge into the filter sump and then flush the chlorine
solution by opening all faucets and let them run until chlorine can no longer be detected (by smell). Your
home/facility has now been completely disinfected and your CALULTRA UV system is ready to inactivate
any microorganisms that enter the water system.
PLEASE NOTE: This procedure must be performed following any time the UV is shut down for service,
without power, or is inoperative for some reason in order to avoid possible contamination of the water
lines.
Cleaning the Quartz Sleeve:
Depending on the water quality, the quartz sleeve may require periodic cleaning. At a minimum, the
quartz sleeve should be cleaned on an annual basis. The following steps outline a basic cleaning
procedure.
Step 1: If the system has an inlet shut-‐off valve, shut this valve off to prevent water flow through the
system. If there is no inlet shut-‐off valve, turn off main water inlet valve (and turn off water pump if you
have one).
Step 2: Disconnect power cord of UV system from electrical outlet.
Step 3: Release water pressure by opening a downstream faucet and close the outlet shut-‐off valve if the
installation incudes one (if there is no outlet shut-‐off valve, be prepared for water leakage from the
system as the head pressure in the system will cause water to flow back through the outlet plumbing and
through the chamber).
Step 4: Remove the captive ground screw from the ground lug on the UV chamber.
Page
12
Summary of Contents for CALULTRA 132
Page 22: ...Page 22 ...
Page 23: ...Page 23 ...