The brief operating instructions in this guide will make the system easier to 

use. As with any surgical instrument, there are important health and safety 

considerations. 

Device Description 
The Benesta

TM

 Tissue Removal Device is a sterile, single-use hand-held 

device that is used to hysteroscopically remove intrauterine tissue. It is 

battery-powered and primarily hand-operated through the use of a button 

that controls the motor inside the handle and the resulting cutting action of 

the blade of the device. Additionally, a safety switch is located on the 

proximal end of the device handle that can be used to turn on or off the 

device. 

Indications for Use 
The Benesta

TM

 Tissue Removal Device is intended for intrauterine use by 

trained gynecologists to hysteroscopically resect and remove tissue such 

as: submucous myomas, endometrial polyps, and retained products of 

conception. 

Contraindications 
The Benesta

TM

 Tissue Removal Device is contraindicated in pregnant 

patients or patients exhibiting pelvic infection, cervical malignancies, or 

previously diagnosed endometrial cancer. 

Warnings 

 

Before using the Benesta

TM

 Tissue Removal Device for the first time, 

please review all available product information. 

 

Before using the Benesta

TM

 Tissue Removal Device, you should be 

experienced in hysteroscopic surgery with powered instruments. Healthy 

uterine tissue can be injured by improper use of the tissue removal 

device. Use every available means to avoid such injury.  

 

Careful pre-operative assessment, including preoperative imaging, 

should be performed on each patient prior to a hysteroscopic procedure 

to evaluate for conditions which may, depending on their severity or 

extent, affect the appropriateness of hysteroscopy. These include but are 

not necessarily limited to: evidence and level of placental invasion of the 

myometrium, acute pelvic inflammatory disease, cervical or vaginal 

infection, known or possible viable pregnancy, carcinoma of the cervix, 

placental invasion of the myometrium, or previously diagnosed 

endometrial cancer.   

 

Removal of retained products of conception in the setting of known or 

suspected placenta accreta, placenta increta, or placenta percreta poses 

a risk of significant and potentially life-threatening bleeding with the 

highest risk occurring in the immediate postpartum phase. 

 

Ensure that a compatible vacuum system that can develop a pressure of 

at least 200 mm Hg is appropriately connected before commencing 

surgery. 

 

Uterine tissue containing suspected fibroids may harbor an occult 

malignancy. The safety of using mechanical tissue removal device has 

not been evaluated in the potential presence of cancer cells. Exercise 

extreme caution when resecting tissue in patients who have implants 

that extend into the uterine cavity. 

– 

Do not use the Benesta

TM

 Tissue Removal Device to resect tissue 

that is adjacent to an implant. When resecting tissue in patients that 

have implants, assure that:  

– 

The Benesta

TM

 Tissue Removal Device cutting window is facing 

away from (i.e., 180° opposite) the implant; 

– 

The visual field is clear; and 

– 

The Benesta

TM

 Tissue Removal Device cutting window is engaged in 

tissue and is moved away from the implant as tissue resection 

proceeds. 

 

If visualization is lost at any point during a procedure, stop cutting 

immediately. 

 

Periodically irrigate the device to prevent accumulation of excised tissue 

in the surgical site. 

 

Operating the device inside the uterine cavity with no tissue contact may 

result in the loss of uterine distension.  

 

Do not use in the presence of flammable or explosive materials. 

 

Not for use in an oxygen-rich environment. 

 

No modification of this equipment is allowed. 

Precautions 

 

Do not use after expiration date. 

 

The Benesta

TM

 Tissue Removal Device is sterilized by ethylene oxide. 

Verify that the Benesta

TM

 Tissue Removal Device is sterile prior to use. 

Do not use the device if the sterile package is open or appears 

compromised. Do not use the device if damage is observed. 

 

The Benesta

TM

 Tissue Removal Device is intended for single use only. 

Do not re-sterilize. Do not reuse. Use of a reprocessed, single-use tissue 

removal device may permanently damage, impede performance, or 

cause failure of the Benesta

TM

 Tissue Removal Device. Use of such 

products may render any warranties null and void. 

 

Discard all opened, unused devices. 

CAUTION: Premature unpacking of the device may result in 

additional and unacceptable risk 

 

Exercise care when inserting or removing the device. Insertion and 

removal of the device should be performed under direct visualization at 

all times. 

 

To avoid perforation, keep the device tip under direct visualization and 

exercise care at all times when maneuvering it or cutting tissue close to 

uterine wall. Avoid using the tip of the tissue removal device as a probe 

or dissecting tool. 

 

Excessive leverage on the Benesta

TM

 Tissue Removal Device does not 

improve cutting performance and, in extreme cases, may result in wear, 

degradation, and seizing of the inner assembly. 

 

Do not allow the cutting window of the tissue removal device to touch 

any metallic object such as a hysteroscope. Damage to both instruments 

is likely. Damage to the Benesta

TM

 Tissue Removal Device can range 

from a slight distortion or dulling of the cutting edge to actual fracture of 

the tip in vivo. If such contact does occur, inspect the tip. If you find 

cracks, fractures, or dulling, or if you have any other reason to suspect a 

tissue removal device is damaged, replace it immediately. 

Summary of Contents for Benesta CAL-TR1511

Page 1: ...ns for Use CAL TR1511 1 pack CAL TR1513 3 pack Manufactured by Caldera Medical Inc 5171 Clareton Drive Agoura Hills CA 91301 U S Toll Free 866 4 CALDERA Telephone 818 879 6555 Fax 818 879 6556 www cal...

Page 2: ...Not Re use Indicates a medical device that is intended for one use or for use on a single patient during a single procedure 1051 5 4 2 Consult Instructions for Use Indicates the need for the user to...

Page 3: ...presence of cancer cells Exercise extreme caution when resecting tissue in patients who have implants that extend into the uterine cavity Do not use the BenestaTM Tissue Removal Device to resect tissu...

Page 4: ...e IV Pole pressure cuff or fluid management system FMS Distending media Vacuum source suction canister and tissue trap BenestaTM Hysteroscope and outflow channel Inflow tubing Y tubing Hysteroscope an...

Page 5: ...ng hard into target pathology Excessive leverage on the device does not improve cutting performance and may result in decreased performance and or unintended excision of tissue pathology 13 Excised ti...

Page 6: ...to earth Not Applicable device is battery operated Radio Frequency RF Common Mode IEC 61000 4 6 3 Vms 150kHz to 80MHz Not Applicable device is battery operated Power Frequency Magnetic Field IEC 61000...

Page 7: ...by Caldera Medical at its option of any product found to be defective during the warranty period Damage inflicted to a product by the user will result in additional charges and may void the warranty T...

Reviews: