background image

 

– 

Connect the outflow vacuum tubing to the tissue trap connection on 

the cannister lid. 

OPERATION 

1.

 

Prior to opening the device, inspect the package for any signs of 

damage. If damage is found, do not use the device and contact 

Technical Support. 

2.

 

Open the device by peeling back the corner of the Tyvek® lid 

completely to expose the device. 

3.

 

Holding the tray securely, grasp the handle of the device firmly and pull 

up to release the device from the tray. 

4.

 

Remove connected outflow tubing and, leaving the end of the tubing 

free, straighten out the tubing so that it is completely unwound. 

5.

 

Turn the safety switch from the “off” (O) position to the “on” ( I ) position. 

6.

 

Temporarily actuate the Benesta

TM

 Tissue Removal Device trigger prior 

to introducing into the working channel of the hysteroscope to ensure 

that the power is on. 

– 

Do not operate the Benesta

TM

 Tissue Removal Device in a non-

irrigated, open-air environment for an extended period of time as this 

may cause damage to the device. 

7.

 

Introduce the inactivated Benesta

TM

 Tissue Removal Device through 

the working channel of the hysteroscope until the distal tip of the tissue 

removal device is visible under clear visualization. 

– 

When used with the Benesta

TM

 Hysteroscope, the tip of the 

Benesta

TM

 Tissue Removal Device will extend approximately 3.2 cm 

from the end of the hysteroscope. 

8.

 

Actuate the Benesta

TM

 Tissue Removal Device trigger for 2-3 seconds 

while under clear visualization and away from any patient tissues to 

prime the device with the distending media. You can also retract the 

tissue removal device into the hysteroscope until the cutting window is 

within the hysteroscope working channel to prime the device. 

– 

If the system is turned off for any reason or the treatment is 

interrupted for any reason, remove the device from any tissue, wait 

15 seconds and repeat steps 5 through 8 before continuing  

9.

 

Rotate to align the cutting window with the target pathology. Rotating 

the hysteroscope may be necessary to achieve desired visualization of 

the cutting window and target tissue pathology. 

10.

 

Bring the cutting window in proximity with the target tissue pathology. 

11.

 

Press the trigger to actuate cutting action in a fashion that allows for 

excision of the excised tissue. 

– 

The suction pressure and/or media infusion pressure can be 

adjusted to optimize visualization and cutting performance. 

– 

Optimal function may be achieved by resecting the tissue in bursts of 

a few seconds each. This may aid in maintenance of distension of 

the endometrial cavity and facilitate monitoring of the cutting 

progress as well as device orientation or position. 

12.

 

Do not apply excessive leverage on the device by pushing hard into 

target pathology. Excessive leverage on the device does not improve 

cutting performance and may result in decreased performance and/or 

unintended excision of tissue pathology. 

13.

 

Excised tissue is collected in the tissue trap. 

Ending the Procedure 
1.

 

Retract the Benesta

TM

 Tissue Removal Device into the hysteroscope 

until the cutting window is within the hysteroscope working channel. 

2.

 

Actuate the trigger for a few seconds to ensure that all tissue is 

transferred into the tissue trap. 

3.

 

Remove the Benesta

TM

 Tissue Removal Device from the hysteroscope. 

Device Disposal 
Dispose of the Benesta

TM

 Tissue Removal Device according to your 

facility’s policies and local, state or Federal procedures for biohazardous 

materials, sharp waste, and battery disposal. 

Storage 

 

The 

Benesta

TM

 Tissue Removal Device should be stored at room 

temperature, away from moisture and direct heat. 

 

The device should not be stored at temperatures below 40°F or above 

120°F, or in conditions where the humidity is below 40% or above 60%. 

Troubleshooting 

 

If the device does not actuate, remove the device from the hysteroscope. 

Turn the safety switch from the “off” (O) position to the “on” ( I ) position. 

Temporarily actuate the trigger to ensure that the power is on before 

reintroducing into the hysteroscope. 

 

If the distal tip of the tissue removal device is not visible after insertion, 

remove the device completely from the hysteroscope. Reintroduce the 

inactivated device through the working channel of the hysteroscope until 

the distal tip of the device is visible under clear visualization. If the issue 

does not resolve, please check the hysteroscope for obstructions and/or 

field of view. Ensure vacuum tubing is connected and functioning and 

vacuum is set to 250mmHg minimum. 

– 

Note: If troubleshooting steps do not resolve the issue, please 

contact Technical Support. 

Electrical and Electromagnetic Safety 

 

All equipment performance is considered to be safety-related 

performance, including any degradation of performance caused by 

reciprocal interference. If the product is not put into service in 

accordance with the electromagnetic safety requirements in this manual 

and fails to perform as specified, the procedure should be aborted 

immediately, and biomedical engineering staff should be alerted to the 

observed issues 

 

The Benesta

TM

 Tissue Removal Device is an applied part Type BF 

device, meaning that it complies with specific requirement for protection 

against electric shock. This device is not suitable for direct cardiac 

application. 

 

The type F applied part status of energized hysteroscopes intended for 

use with a multiplicity of supply units and/or light guide cables is ensured 

by, for instance, using only supply units having isolated light guide output 

sockets 

 

The standard operating voltage of the Benesta

TM

 Tissue Removal Device 

is 7.7V – 9.0V, and the operating current is ~0.2A. 

Summary of Contents for Benesta CAL-TR1511

Page 1: ...ns for Use CAL TR1511 1 pack CAL TR1513 3 pack Manufactured by Caldera Medical Inc 5171 Clareton Drive Agoura Hills CA 91301 U S Toll Free 866 4 CALDERA Telephone 818 879 6555 Fax 818 879 6556 www cal...

Page 2: ...Not Re use Indicates a medical device that is intended for one use or for use on a single patient during a single procedure 1051 5 4 2 Consult Instructions for Use Indicates the need for the user to...

Page 3: ...presence of cancer cells Exercise extreme caution when resecting tissue in patients who have implants that extend into the uterine cavity Do not use the BenestaTM Tissue Removal Device to resect tissu...

Page 4: ...e IV Pole pressure cuff or fluid management system FMS Distending media Vacuum source suction canister and tissue trap BenestaTM Hysteroscope and outflow channel Inflow tubing Y tubing Hysteroscope an...

Page 5: ...ng hard into target pathology Excessive leverage on the device does not improve cutting performance and may result in decreased performance and or unintended excision of tissue pathology 13 Excised ti...

Page 6: ...to earth Not Applicable device is battery operated Radio Frequency RF Common Mode IEC 61000 4 6 3 Vms 150kHz to 80MHz Not Applicable device is battery operated Power Frequency Magnetic Field IEC 61000...

Page 7: ...by Caldera Medical at its option of any product found to be defective during the warranty period Damage inflicted to a product by the user will result in additional charges and may void the warranty T...

Reviews: