background image

 

 

Exercise care when inserting or removing the device. Excessive bending 

of the device distal tip can cause the Benesta

TM

 Tissue Removal Device 

cutter to come out of the cutting window. If such damage occurs, replace 

the device immediately. 

Benesta

TM

 Tissue Removal Device 

  CAL-TR1511 (1 pack) 

  CAL-TR1513 (3 pack) 

 

The hand-operated Benesta

TM

 Tissue Removal Device is shown below in 

Figure 1. The parts of the device are listed below: 

 

Figure 1. Benesta

TM

 Tissue Removal Device  

Safety Switch: This switch turns power to the entire device on or off 

depending on the position of the switch. During shipment, the position is 

in the “off” (O) position to prevent any accidental activation. The safety 

switch must be turned to the “on” (I) during the preparation phase of an 

operative procedure for the device to function and resect tissue. 
Trigger: Depressing the trigger actuates cutting action. The device will 

continuously cut and allow the passage of fluid and tissue while the 

trigger is depressed and will stop as soon as the trigger is released. The 

cutting window will remain closed while cutting is not activated, but will 

still allow for some fluid outflow in this position. 
Outflow Tubing: the outflow tube removes waste fluid and resected 

tissue through a compatible suction canister and tissue trap. 
Cutting Window: side-facing cutting window at the distal tip 
Cutting Window Indicator: indicates orientation of cutting window when 

cutting window is not under direct visualization 
Distal Tip Locating Mark: Indicates depth relative to hysteroscope and 

can be used during initial insertion of the tissue removal device into the 

hysteroscope. 
Pathology Measurement Lines: contains markings spaced 1mm apart to 

help estimate pathology size under direct visualization 

 

Components not included with the Benesta

TM

 Tissue Removal Device: 

– 

Under-buttocks drape 

– 

IV Pole, pressure cuff or fluid management system (FMS)  

– 

Distending media 

– 

Vacuum source, suction canister and tissue trap 

– 

Benesta

TM

 Hysteroscope and outflow channel 

– 

Inflow tubing 

– 

Y-tubing 

– 

Hysteroscope and outflow channel end cap seals 

SET-UP 

Please review the system configuration diagram in Figure 2 for a diagram 

of the system setup. 

 

Figure 2. System configuration diagram 

Distending Media Set-Up 

 

It is strongly recommended that isotonic media such as 0.9% Normal 

Saline or Ringer’s Lactate be the media used for distending the 

endometrial cavity. 

 

Either a purpose-built hysteroscopic fluid management system or a 

protocol for manual estimation of systemic absorption should be in place. 

 

For purpose-built fluid management systems, carefully follow the 

manufacturer’s instructions. 

 

If using a manual fluid management system: 

– 

Procure an intravenous (IV) pole from which to hang either a 1L or 

3L bag of distending media at least 40 inches above the patient’s 

uterus 

– 

A pressure sleeve may be placed around the fluid bag to increase 

inflow and uterine pressure and improve visualization during the 

procedure. 

– 

When using the Benesta

TM

 Hysteroscope, open the packaging for the 

Y-tubing and connect it to the hysteroscope inflow ports. Connect the 

Y-tubing to the fluid bag containing distention media using 

compatible inflow tubing. 

– 

When using with the Benesta

TM

 Hysteroscope (and/or outflow 

channel), make sure the single-use seal is installed in the 

hysteroscope (and/or outflow channel). 

 

Position an under-buttocks drape with an integrated fluid reservoir or 

similar fluid collection system that allows estimation of fluid volume. 

 

Connect the outflow tubing to a tissue trap in line with tubing that 

attaches to an appropriate suction canister. 

 

Attach the suction canister to the regulated vacuum source which is set 

to approximately 250 mmHg. 

– 

Make sure the lid is securely attached to the cannister prior to using.  

Safety 

Switch 

Outflow 

Tubing

 

 

Trigger

 

Distal Tip 

Locating Mark

 

 

 

Cutting Window

 

 

Cutting Window Indicator

 

Pathology Measurement. Lines

 

Summary of Contents for Benesta CAL-TR1511

Page 1: ...ns for Use CAL TR1511 1 pack CAL TR1513 3 pack Manufactured by Caldera Medical Inc 5171 Clareton Drive Agoura Hills CA 91301 U S Toll Free 866 4 CALDERA Telephone 818 879 6555 Fax 818 879 6556 www cal...

Page 2: ...Not Re use Indicates a medical device that is intended for one use or for use on a single patient during a single procedure 1051 5 4 2 Consult Instructions for Use Indicates the need for the user to...

Page 3: ...presence of cancer cells Exercise extreme caution when resecting tissue in patients who have implants that extend into the uterine cavity Do not use the BenestaTM Tissue Removal Device to resect tissu...

Page 4: ...e IV Pole pressure cuff or fluid management system FMS Distending media Vacuum source suction canister and tissue trap BenestaTM Hysteroscope and outflow channel Inflow tubing Y tubing Hysteroscope an...

Page 5: ...ng hard into target pathology Excessive leverage on the device does not improve cutting performance and may result in decreased performance and or unintended excision of tissue pathology 13 Excised ti...

Page 6: ...to earth Not Applicable device is battery operated Radio Frequency RF Common Mode IEC 61000 4 6 3 Vms 150kHz to 80MHz Not Applicable device is battery operated Power Frequency Magnetic Field IEC 61000...

Page 7: ...by Caldera Medical at its option of any product found to be defective during the warranty period Damage inflicted to a product by the user will result in additional charges and may void the warranty T...

Reviews: