Portable Oxygen Concentrator
3.
Remove the compressor by pulling the compressor away from the PSA assembly.
Care should be taken not to separate the manifold from the beds.
Figure 14.2: Removing the Compressor
4.
If either of the two fittings connecting the compressor to the manifold is connected to the
compressor, remove it and insert it back into the manifold.
Figure 14.3: Compressor to Manifold Fittings
5.
Connect the new compressor to the PSA assembly, and tighten the bolt to a torque of 15
in-lbs.