MN220-1 B 78
Portable Oxygen Concentrator
Figure 22.2: Disconnecting Flex Circuit
6.
Turn the unit over so that the main circuit board is facing upwards.
7.
Remove the motor controller board per the MOTOR CONTROLLER BOARD REPLACEMENT
section of this manual.
8.
If replacing the circuit board on a FreeStyle, remove the hour meter per the HOUR METER
REPLACEMENT section of this manual.
9.
Remove the cannula nozzle tuning by cutting the tie wrap (use small-size cutter). Carefully
pull the tubing from the nozzle barb.