MN220-1 B 10
Portable Oxygen Concentrator
FREESTYLE AND FREESTYLE 5 ALARMS
Status
Audible Alarm
Light
Indicates
Action
Start-Up
Brief, continuous
at start–up
(Green) pulse;
continuous light
FreeStyle or
FreeStyle 5 has been
turned on
You may begin to operate your
unit.
Normal
Operation
No
(Green) pulse;
intermittent light
FreeStyle or
FreeStyle 5 is
delivering oxygen as
a pulse flow.
Continue using unit normally.
Alarm
Continuous:
Beep
(Red) alarm;
continuous light
No breath detected
by the unit for a
predetermined time
period.
Check the cannula connection.
Ensure that you are breathing
through your nose. If the
alarm persists, contact your
Equipment provider.
Alarm
Intermittent:
Beep, beep, beep.
25% (Yellow)
BATT; battery
indicator.
Battery voltage is too
low to operate the
FreeStyle or
FreeStyle 5
Connect the FreeStyle or
FreeStyle 5 unit into a DC
outlet or an AC outlet
immediately.
Alarm
Rapid intermittent:
Beep, beep, beep
(Red) alarm;
intermittent light
Breathing rate is
exceeding the
capacity of the
FreeStyle or
FreeStyle 5 unit.
Reduce activity, then if
necessary use another source
of oxygen as available. Contact
your Equipment Provider.
Alarm
Rapid intermittent:
Beep, beep, beep
(Red) alarm;
continuous light
General malfunction
of the FreeStyle or
FreeStyle 5 unit has
occurred.
Turn off the unit. Change to
another source of oxygen, and
contact your